• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ESSENTIO MRI SR; IMPLANTABLE PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ESSENTIO MRI SR; IMPLANTABLE PACEMAKER Back to Search Results
Model Number L110
Medical Device Problem Code Nonstandard Device (1420)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/30/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that this pacemaker was received by post market quality assurance without a known reason for explant.No adverse patient effects were reported.
 
Additional Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, this device was thoroughly inspected and analyzed.A high current drain was detected.Detailed analysis confirmed the identified high current drain was a result of compromised capacitors.This resulted in the observed premature battery depletion.It was determined that the capacitors were compromised due to the presence of excess hydrogen in the device case.Boston scientific issued a field safety notice regarding a subset of devices in the accolade pacemaker family that have a potential of exhibiting this behavior.This particular device is included in the hydrogen induced premature depletion advisory population.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ESSENTIO MRI SR
Common Device Name
IMPLANTABLE PACEMAKER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112 5798
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key24536521
Report Number2124215-2026-12690
Device Sequence Number8895523
Product Code LWP
UDI-Device Identifier00802526558948
UDI-Public(01)00802526558948(17)180614
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
P150012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Remedial Action Notification,Recall
Type of Report Initial
Report Date (Section B) 03/06/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date06/14/2018
Device Model NumberL110
Device Catalogue NumberL110
Device Lot Number718863
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/30/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/20/2026
Initial Report FDA Received Date03/06/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/27/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0367-2018
Patient Sequence Number1
Patient Age74 YR
Patient SexFemale
-
-