BOSTON SCIENTIFIC CORPORATION EASYTRAK 2 IS-1; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D Back to Search Results
Model Number 4542
Medical Device Problem Codes Use of Device Problem (1670); Connection Problem (2900)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/16/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If pertinent information is provided in the future, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that interrogation of this cardiac resynchronization therapy defibrillator (crt-d) system was unsuccessful, and this crt-d had reached the explant indicator in (b)(6) 2024.Review of device settings noted low out-of-range pace impedance measurements less than 200 ohms for more than a year on the right ventricular (rv) lead, which was purposefully connected to the left ventricular (lv) port at the last device replacement procedure.Additional information received reported that the rv lead impedance issue was known at the previous device changeout, and the rv and lv leads were purposefully switched in the header during the battery replacement procedure for this reason.At this time, there are no other plans for intervention, other than the upcoming battery replacement.This lv lead remains in service.No additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
If pertinent information is provided in the future, a supplemental report will be submitted.This product was manufactured prior to implementation of the udi regulation and as a result, the complete udi is not available.Applicable us product data has been included in this report.Correction to b2: populated outcomes attrib to adv event.
 
Event or Problem Description
It was reported that interrogation of this cardiac resynchronization therapy defibrillator (crt-d) system was unsuccessful, and this crt-d had reached the explant indicator in october 2024.Review of device settings noted low out-of-range pace impedance measurements less than 200 ohms for more than a year on the right ventricular (rv) lead, which was purposefully connected to the left ventricular (lv) port at the last device replacement procedure.Additional information received reported that the rv lead impedance issue was known at the previous device changeout, and the rv and lv leads were purposefully switched in the header during the battery replacement procedure for this reason.At this time, there are no other plans for intervention, other than the upcoming battery replacement.This lv lead remains in service.No additional adverse patient effects were reported.
 
Event or Problem Description
It was reported that interrogation of this cardiac resynchronization therapy defibrillator (crt-d) system was unsuccessful, and this crt-d had reached the explant indicator in october 2024.Review of device settings noted low out-of-range pace impedance measurements less than 200 ohms for more than a year on the right ventricular (rv) lead, which was purposefully connected to the left ventricular (lv) port at the last device replacement procedure.Additional information received reported that the rv lead impedance issue was known at the previous device changeout, and the rv and lv leads were purposefully switched in the header during the battery replacement procedure for this reason.At this time, there are no other plans for intervention, other than the upcoming battery replacement.This lv lead remains in service.No additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
If pertinent information is provided in the future, a supplemental report will be submitted.This product was manufactured prior to implementation of the udi regulation and as a result, the complete udi is not available.Applicable us product data has been included in this report.Correction to b2: populated outcomes attrib to adv event.This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.A review of the device history record (dhr) was performed.The review of the dhr identified that there were no process related non-conformances, scrap, or rework performed during the production that could explain the event.The reviews ensure each device meets specification prior to release for use.There is no indication the device manufacturing process contributed to the reported complaint.Review of labeling determined that the complaint situation was listed in the manual.The manual was unlikely to be the cause of the reported complaint; translation, wording, or graphics does not require further review.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of unintended use error caused or contributed to event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EASYTRAK 2 IS-1
Common Device Name
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
CPI - DEL CARIBE
guidant puerto rico b. v.
#12, rd. #698
dorado PR 00646
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key24537246
Report Number2124215-2026-12713
Device Sequence Number15710084
Product Code NIK
Combination Product (Y/N)Y
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P010012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Remedial Action Notification,Recall
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/16/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date02/14/2010
Device Model Number4542
Device Catalogue Number4542
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/16/2018
Supplement Date Received by Manufacturer03/09/2026
06/15/2026
Initial Report FDA Received Date03/06/2026
Supplement Report FDA Received Date03/11/2026
06/16/2026
Date Device Manufactured02/14/2008
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age71 YR
Patient SexFemale
-
-