INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD MODIFIED SKULL CLAMP Back to Search Results
Catalog Number A1059
Medical Device Problem Code Device Slipped (1584)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 02/10/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event or Problem Description
A facility reported that mayfield modified skull clamp (a1059) slipped while in use; however, they are unsure if it was user or product error.It is unknown if there was injury or delay.
 
Additional Manufacturer Narrative
Updated fields: d4, d9, g3, g6, h2, h3, h4, h6, h11.The mayfield modified skull clamp (a1059) was not returned; however, lot number information has been provided; therefore, device history record (dhr) was reviewed, and no anomalies were found.Failure analysis cannot be completed due to the lack of information received to perform a complete investigation; therefore, the definite root cause cannot be determined.Based on the reported complaint, the probable root cause is improper or suboptimal placement of the skull clamp on the patient.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
 
Event or Problem Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD MODIFIED SKULL CLAMP
Common Device Name
MAYFIELD
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH 45227
Manufacturer (Section G)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
cincinnati OH 45227
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 08540
6099362319
MDR Report Key24539305
Report Number3004608878-2026-00023
Device Sequence Number21430973
Product Code HBL
UDI-Device Identifier10381780253457
UDI-Public10381780253457
Combination Product (Y/N)N
PMA/510(K) Number
PRE-AMEND
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberA1059
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/10/2026
Supplement Date Received by Manufacturer04/14/2026
Initial Report FDA Received Date03/06/2026
Supplement Report FDA Received Date04/24/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/18/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-