COCHLEAR BONE ANCHORED SOLUTIONS AB BIA400 IMPLANT 4MM W ABUTMENT 8MM; COCHLEAR BAHA CONNECT SYSTEM Back to Search Results
Model Number 93330
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Codes Impaired Healing (2378); Skin Infection (4544)
Type of Reportable Event Serious Injury
Event or Problem Description
Per the clinic, the patient experienced infection at the abutment site with associated tissue overgrowth around the site.There are plans to explant the device and re-implant the patient with a new device; however, this has not occurred as of the date of this report.Additional information has been requested but has not been made available as of the date of this report.
 
Additional Manufacturer Narrative
Per the clinic, the device was explanted under general anesthesia on (b)(6) 2026.The reported adverse event is associated with a returned device; however, the provided clinical information was reviewed by the manufacturer and no specific device analysis is deemed necessary at this time.Previous product examinations have not showed any relationship between a product geometrical deviation and the reported clinical complication.Additionally, there are no indications that a product failure has contributed to the reported issue.Adverse skin reactions around the baha abutment, ranging from slight redness to infected tissue, are common complications with baha implants, and can occur at any time following implantation.The report frequency for these complications is being monitored under cbas complaint and medical device reporting data monitoring plan and the status is updated on a monthly basis.This event is added to this monitoring.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIA400 IMPLANT 4MM W ABUTMENT 8MM
Common Device Name
COCHLEAR BAHA CONNECT SYSTEM
Manufacturer (Section D)
COCHLEAR BONE ANCHORED SOLUTIONS AB
konstruktionsvägen 14
mölnlycke, 435 3 3
SW  435 33
Manufacturer (Section G)
COCHLEAR BONE ANCHORED SOLUTIONS AB
konstruktionsvägen 14
mölnlycke, 435 3 3
SW   435 33
Manufacturer Contact
amirah rosly
unit ug-1, vertical podium
no. 8 jalan kerinchi
kuala lumpur, wilayah persekutuan 59200
MY   59200
MDR Report Key24542257
Report Number6000034-2026-00865
Device Sequence Number5573127
Product Code MAH
UDI-Device Identifier09321502022705
UDI-Public(01)09321502022705(10)166359(17)191031
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K121317
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2026
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/08/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number93330
Device Catalogue Number93330
Device Lot Number166359
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/11/2026
Supplement Date Received by Manufacturer03/12/2026
Initial Report FDA Received Date03/08/2026
Supplement Report FDA Received Date04/08/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-