• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON LABORATORIES IRELAND LTD. CLAREON ASPHERIC UV ABSORBING IOL WITH AUTONOME DELIVERY SYSTEM; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON LABORATORIES IRELAND LTD. CLAREON ASPHERIC UV ABSORBING IOL WITH AUTONOME DELIVERY SYSTEM; INTRAOCULAR LENS Back to Search Results
Model Number CCA0T0
Medical Device Problem Codes Device Difficult to Setup or Prepare (1487); Failure to Unfold or Unwrap (1669)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/11/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A facility representative reported during intraocular lens (iol) implant procedure, the surgeon noticed the while being loaded haptic was folded under the optic.The procedure was completed on same day.There was no patient contact.Additional information was requested, yet no new information was received.
 
Additional Manufacturer Narrative
Correction information was updated (a040604 was updated to a2201) additional information was provided in d.9., h.3., h.6., and h.11.The product was returned for analysis, and the reported complaint was not observed.Straight leading haptic was observed.The device was received loose in the product carton.Solution is dried in the device.The lock-out assembly has been removed.The plunger and the lens have been advanced into the mid-nozzle.The plunger position is acceptable.The trailing haptic is folded onto the optic.The leading haptic is extended straight.The reported "haptic folded under the optic" was not observed.No root cause identified as the reported "haptic folded under the optic" was not observed.Straight leading haptic was observed.The presentation of straight-leading haptics is addressed in the product instructions for use (ifu), where it is provided that delivery may proceed with caution and should be managed based on the outlined instructions.While straight leading haptics may occasionally occur, close adherence to the ifu provides the best opportunity for optimal delivery.Possible associated factors may also include: use of a non-qualified viscoelastic or bss in the delivery system.Material properties of non-qualified ovds (ophthalmic viscosurgical device) may contribute to underfill, overfill, misfolding of the haptics, or other inconsistent folding outcomes.Excessive delay between device preparation and subsequent delivery resulting in the potential for haptic unfolding.Use of the delivery system at operating room temperatures outside of the recommended range of 18 °c (64 °f) to 23 °c (73 °f).Ovd should be allowed to come to operating room temperature over a 20-minute timeframe prior to use.Any of the above factors alone, or in combination, may lead to an event such as reported.Based on the results from the product history record, the products met release criteria.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON ASPHERIC UV ABSORBING IOL WITH AUTONOME DELIVERY SYSTEM
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI  00000
Manufacturer (Section G)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI   00000
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key24544724
Report Number9612169-2026-00527
Device Sequence Number5573197
Product Code HQL
UDI-Device Identifier00380652397353
UDI-Public(01)00380652397353(17)291105(11)251107(10)26144220
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/01/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCCA0T0
Device Lot Number26144220
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/13/2026
Supplement Date Received by Manufacturer05/07/2026
Initial Report FDA Received Date03/09/2026
Supplement Report FDA Received Date06/01/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/07/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKNOWN VISCOELASTIC
Patient SexUnknown
-
-