| Model Number |
E1455NSB |
| Medical Device Problem Codes |
Peeled/Delaminated (1454); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Codes |
Unintended Radiation Exposure (4565); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/10/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that during use, the blue plastic protective covering on the electrode blade slid off.The covering was immediately identified, retrieved, and replaced.There were no device components retained in the patient.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
D10 concomitant product: e1455nsb, e1455nsb b the edge ent/ima electr x50 (lot#252860205r).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: b1, b2, b5, d9, e1 (initial reporter), e2, e3, e4 (email), g3, h1, h3, h6 (ime) this event has been reassessed and the reportability has been determined to be a reportable serious injury.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that during use, the blue plastic protective covering on the electrode blade slid off.The covering was immediately identified, an x-ray was performed to retrieved the piece and there were no device components retained in the patient.There were no additional medical procedure needed.The covering was then replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device and a photo were available for evaluation.Visual inspection noted that the blue shrink wrap was also observed.Clear piece of insulation is no longer attached snug.Functional testing found that when attaching and detaching the electrode, user should be grasping the electrode by the more durable overmold.Using tools will inflict damage and may result in detachment of the clear piece of insulation.It was reported that the device component was disengaged and the insulation was damaged.The reported issues were confirmed.The product analysis noted evidence that the device was not used as intended.This issue may occur with consistent misuse.Off label technique inserting and extracting the electrode and/or modifying the insulation.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.The instructions included with this device provide the following guidance: do not modify or add to the insulation of activate electrodes.Grasp the insulation sleeve on the electrode.Insert the electrode into the pencil.Inspect the electrode for insulation damage and coating damage.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|