Visual inspection: during the investigation, scratches on the outer housing of the valve were determined.Permeability test: a permeability test has shown that the m.Blue is permeable.Computer controlled test: to investigate the claim of over- drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valve is tested in the horizontal position.The results show that the valve operates within the accepted tolerances in both positions.Adjustment test: the m.Blue valve was tested and is adjustable to all specified pressures.Braking force and brake function test: the brake functionality test has shown that the brake function is fully operational, and the braking force is within the given tolerances.Internal inspection: after dismantling of the valve, no deposits were found in m.Blue results: based on our investigation results, we cannot determine any functional deviations or other abnormalities in the product.The valve is operating within specifications.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
|