MAUDE Adverse Event Report: ABBOTT MEDICAL TACTICATH¿ CONTACT FORCE ABLATION CATHETER, SENSOR ENABLED¿ BID CURVE D-F; Catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL TACTICATH¿ CONTACT FORCE ABLATION CATHETER, SENSOR ENABLED¿ BID CURVE D-F; Catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation Back to Search Results
Model Number A-TCSE-DF
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code Cardiac Perforation (2513)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Review of the device history record was not possible as the lot number is unknown.Based on the information received, the cause of the reported incident could not be conclusively determined.Per the ifu, cardiac tamponade and pericardial effusion are known risks during the use of this device.The product's lot number could not be obtained, therefore the primary di number is provided (d4), but full udi information is not available.
 
Event or Problem Description
An baifu, zhang xuelian, bai mei, et al.Effect of using impedance change monitoring to assist lesion index¿guided high-power ablation for atrial fibrillation.Chinese journal of geriatrics, 2025, 45(08): 1793¿1796.A total of 50 patients who underwent high-power (40¿45 w) atrial fibrillation ablation guided by the lesion size index (lsi) were enrolled as the control group.Another 50 newly selected patients who met the indications for radiofrequency ablation of atrial fibrillation were included as the study group.Ablation in the study group was initiated using the same high-power settings as the control group.During the procedure, the baseline impedance and impedance drop at each ablation point were monitored.When the lsi reached the target value but the impedance drop was <10%, the ablation parameters and the contact force of the ablation catheter were adjusted, followed by additional ablation.If an impedance rise occurred during ablation or if the impedance dropped by >15% within the first 5 seconds, the ablation was immediately terminated.4 steam pops and 1 steam pop followed by a pericardial effusion requiring a pericardiocentesis were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TACTICATH¿ CONTACT FORCE ABLATION CATHETER, SENSOR ENABLED¿ BID CURVE D-F
Common Device Name
Catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation
Manufacturer (Section D)
ABBOTT MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ABBOTT MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
wendy mccarty
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key24547008
Report Number3008452825-2026-00126
Device Sequence Number12193816
Product Code OAE
UDI-Device Identifier05415067027641
UDI-Public05415067027641
Combination Product (Y/N)N
Initial Reporter CountryCH
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Literature
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/09/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberA-TCSE-DF
Device Catalogue NumberA-TCSE-DF
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/12/2026
Initial Report FDA Received Date03/09/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening; Required Intervention;
Patient SexUnknown
-
-