| Model Number |
10007 |
| Medical Device Problem Code |
Failure to Deliver (2338)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/27/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A delivery failure alarm due to a main supply voltage below tolerance was identified by a internal employee during a routine service log review.Further review of the service log determined that a low battery alarm did not occur prior to the delivery failure alarm.As a result, these service log findings meet the criteria of a reportable malfunction as it was located at a user's facility and set to administer a dose of inhaled nitric oxide (no) at the time of the incident.This device was located in the united states when the reportable malfunction occurred.There was no delivery failure alarm complaint or report of patient harm associated with this incident.
|
| |
|
Additional Manufacturer Narrative
|
|
Reportable malfunction with inomax dsir plus (b)(6) was documented and investigated under (b)(4).A review of the device's service log revealed that a delivery failure alarm occurred due to the main supply voltage being out of tolerance (valid range: 2435 to 4075 counts, actual value: 2431 counts).Further review of the device's service log determined that a low battery alarm did not occur to prompt the user to plug the device into a power source.A low battery alarm should have occurred prior to the delivery failure alarm to prevent the out of tolerance main supply voltage condition.The regional servicecenter (rsc) did not experience a delivery failure alarm without a low battery alarm during testing of the returned device on battery power.The root cause for this occurrence was determined to be battery fuel gauge drift.As a result, the device's battery was replaced.Trends were reviewed for this reportable condition and determined to be within the established control limits.No further action is required at this time.A full functional test was performed and the device operated according to specifications.As a result, the device was placed back into circulation for customer use.
|
| |
|
Search Alerts/Recalls
|