| Model Number |
PM1240 |
| Medical Device Problem Code |
Premature Discharge of Battery (1057)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/12/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Following the electronics performance indicator (epi), an alert was received by the clinician and awaiting device explant. further information was requested, but not yet available.The patient condition was not reported.
|
| |
|
Event or Problem Description
|
|
New information indicates that the pacemaker was explanted replaced on (b)(6) 2026.The patient condition was stable.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event of premature discharge of battery was not confirmed.The device was above elective replacement indicator (eri) level upon receipt.Visual inspection of the header found no anomaly related to the field concern.Review of the device image indicates auto-capture was enabled, consistent with electronics performance indicator (epi) triggered.When automatic settings are programmed, such as auto-capture, the device output would adjust itself to accommodate the patient¿s needs for pacing.Electrical and mechanical analysis performed indicated normal functionality.Further battery assessment at various pulse amplitudes found normal current levels and no indication of premature battery depletion noted.Longevity assessment was performed, and the device was in normal range of operation with appropriate remaining longevity.
|
| |
|