| Model Number |
EMAX-25 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Atrial Fibrillation (1729); Tachycardia (2095)
|
| Date of Event |
12/13/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device remains in patient.The device will not be returned for evaluation.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
Clinical information: (b)(4) - sh device registry, patient site id: (b)(6).It was reported that on (b)(6) 2025, a 25mm epic max valve was implanted in the patient.After the procedure, an atrial fibrillation and r-on-t phenomenon that develops into torsades de pointes (tdp) tachycardia had occurred.On (b)(6) 2026, a defibrillation was performed.On (b)(6) 2026, an implantable cardioverter defibrillator (icd) was implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of arrhythmia was reported.A returned device assessment could not be performed as the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed, and the product met all specifications.Based on the available information, the cause of the reported arrhythmia could not be conclusively determined.It is possible that procedural circumstances could have contributed to the reported event.The reported unexpected medical intervention and surgical intervention were results of case-specific circumstances.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.
|
| |
|
Event or Problem Description
|
|
Clinical information: (b)(4) - sh device registry, patient site id: (b)(6).It was reported that on (b)(6) 2025, a 25mm epic max valve was implanted in the patient.After the procedure, an atrial fibrillation and r-on-t phenomenon that develops into torsades de pointes (tdp) tachycardia had occurred.On (b)(6) 2026, a defibrillation was performed.On (b)(6) 2026, an implantable cardioverter defibrillator (icd) was implanted.The patient was reported discharged.
|
| |
|