| Catalog Number |
528135 |
| Medical Device Problem Code |
Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed, and no relevant findings were identified.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.If the sample becomes available at a later date a follow-up report will be submitted with investigation results.Teleflex will continue to monitor and trend for reports of this nature.
|
| |
|
Event or Problem Description
|
|
It was reported that "staple jamming." a new stapler was used to proceed with the procedure.No patient harm or injury reported.
|
| |
|
Event or Problem Description
|
|
It was reported that "staple jamming." a new stapler was used to proceed with the procedure.No patient harm or injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The customer returned one unit of 528135 visistat 35r 6/box for investigation.The returned stapler was visually examined with and w ithout magnification.Visual examination of the returned stapler revealed that the staples appeared properly aligned.The stapler was returned with the trigger not engaged, and there were signs of biological material on the device.The stapler was received with 3 staples left in the cartridge, indicating at least 32 staples were fired by the end user.A functional inspection was performed on the sample by attempting to fire staples from the returned stapler.Upon full engagement of the trigger, the first staple fully formed and released in the air.To simulate insertion into the skin, the stapler was fired into a simulated skin pad.The remaining staples were fired and were able to engage and close into the simulated skin.No functional problem was found with the returned sample.The ifu for this product was reviewed as a part of this complaint investigation.The ifu states, "to obtain optimum staple closure, the trigger must be squeezed all the way in." a corrective action is not required at this time, as there were no functional issues found with the returned sample.The reported complaint of "misfire/jamming - info not provided" could not be confirmed based upon the sample received.The customer returned one unit of 528135 visistat 35r 6/box for investigation.The returned stapler was able to form and release all remaining staples in the cover block.A device history record review was performed on the device with no evidence to suggest a manufacturing-related root cause.There were no functional issues found with the returned stapler.Teleflex will continue to monitor and trend on complaints of this nature.Other remarks: n/a.Corrected data: n/a.
|
| |
|