Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/29/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H.3.A follow up mdr will be submitted if additional information, a device evaluation, or a device history review is completed.
|
| |
|
Event or Problem Description
|
|
When the j-loop was connected to the 24-gage catheter it leaked when flushed.Additional information 3/4/26: any adverse event or serious injury reported to patient or healthcare professional? if yes, please provide the details.None known.Can you please provide an exact date of event? 01-29-2026 we had a second event as well, so i am not 100% sure which one this is referencing.The other event date is 10-20-2025.
|
| |
|
Event or Problem Description
|
|
No new information.
|
| |
|
Additional Manufacturer Narrative
|
|
A device history record review was completed by our quality engineer team for provided material number 381412 and lot number 5226488.The review did not reveal any detected abnormalities during the production process that could have contributed to the reported issue, and all quality tests were found to be within specification.As a sample was unavailable for return, a thorough sample investigation could not be completed.Based on the investigation results, an exact cause for this incident could not be identified.There are quality controls currently in place to detect similar issues during the production process.Further action has not been determined necessary at this time.Complaints received for this device and reported condition will continue to be tracked and trended.Our quality team regularly reviews the collected data for identification of emerging trends.
|
| |
|
Search Alerts/Recalls
|