| Model Number |
33151 |
| Medical Device Problem Codes |
Use of Device Problem (1670); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
Bowel Burn (1756)
|
| Date of Event |
02/12/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that during laparoscopic surgery for the treatment of endometriosis, a unipolar high frequency cord was attached to the handle for use with a maryland grasper.The metal jaws were applied to tissue deep within the utero pelvic region, and the surgeons ensured that no surrounding structures were involved before applying coagulation.Cross reference mdr: 9610617-2026-00500.9610617-2026-00501.
|
| |
|
Additional Manufacturer Narrative
|
|
The affected device has been requested for investigation by the manufacturer.The device was not yet returned for investigation.The event is filed under internal karl storz complaint id: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The device in question was returned to the manufacturer.The evaluation is anticipated, but not yet begun.The investigation will be performed by a designated karl storz employee.The event is filed under internal karl storz complaint id: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation:
all returned items were carefully inspected, and no damage was identified on any of them beyond normal wear and tear.Based on the customer's description, an electrical current was allegedly detected in the upper section near the handle.However, the handle is made entirely of plastic, which makes it impossible for electrical current to pass through it.The only plausible explanation would be damage to the insulation of the shaft, allowing current to leak and potentially cause a burn.Unfortunately, we do not have any images, information on the age, or the exact model number of the outer shaft.Despite multiple follow-up requests, no additional details could be obtained.No corrective actions necessary as there is no critical and/or systematic deviation found during investigation.The event is filed under internal karl storz complaint id: (b)(4).
|
| |
|
Search Alerts/Recalls
|