• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KARL STORZ SE & CO. KG CLICKLINE FORCEPS INSERT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KARL STORZ SE & CO. KG CLICKLINE FORCEPS INSERT Back to Search Results
Model Number 33310ML
Medical Device Problem Codes Use of Device Problem (1670); Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Bowel Burn (1756)
Date of Event 02/12/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The affected device has been requested for investigation by the manufacturer.The device was not yet returned for investigation.The event is filed under internal karl storz complaint id: (b)(4).
 
Event or Problem Description
It was reported that during laparoscopic surgery for the treatment of endometriosis, a unipolar high frequency cord was attached to the handle for use with a maryland grasper.The metal jaws were applied to tissue deep within the utero pelvic region, and the surgeons ensured that no surrounding structures were involved before applying coagulation.Cross reference mdr: 9610617-2026-00499; 9610617-2026-00500.
 
Additional Manufacturer Narrative
The device in question was returned to the manufacturer.The evaluation is anticipated, but not yet begun.The investigation will be performed by a designated karl storz employee.The event is filed under internal karl storz complaint id: (b)(4).
 
Additional Manufacturer Narrative
Investigation: all returned items were carefully inspected, and no damage was identified on any of them beyond normal wear and tear.Based on the customer's description, an electrical current was allegedly detected in the upper section near the handle.However, the handle is made entirely of plastic, which makes it impossible for electrical current to pass through it.The only plausible explanation would be damage to the insulation of the shaft, allowing current to leak and potentially cause a burn.Unfortunately, we do not have any images, information on the age, or the exact model number of the outer shaft.Despite multiple follow-up requests, no additional details could be obtained.No corrective actions necessary as there is no critical and/or systematic deviation found during investigation.The event is filed under internal karl storz complaint id: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLICKLINE FORCEPS INSERT
Common Device Name
CLICKLINE FORCEPS INSERT
Manufacturer (Section D)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM  78532
Manufacturer (Section G)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM   78532
Manufacturer Contact
anja fair
2151 e grand ave
el segundo, CA 90245
MDR Report Key24553363
Report Number9610617-2026-00501
Device Sequence Number12156003
Product Code GCJ
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K935071
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 06/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number33310ML
Device Catalogue Number33310ML
Device Lot NumberVK45
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/13/2026
Supplement Date Received by Manufacturer03/18/2026
05/27/2026
06/17/2026
Initial Report FDA Received Date03/09/2026
Supplement Report FDA Received Date03/25/2026
06/08/2026
06/17/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-