• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOSENSE WEBSTER INC QDOT MICRO; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOSENSE WEBSTER INC QDOT MICRO; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION Back to Search Results
Catalog Number D139505
Medical Device Problem Codes Incorrect, Inadequate or Imprecise Result or Readings (1535); Contamination /Decontamination Problem (2895)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/24/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device evaluation details: the qdot micro device was returned to johnson & johnson medtech (j&j medtech) for evaluation.A visual inspection and screening test of the returned device were performed following j&j medtech procedures.Visual analysis revealed a reddish material and a hole in the surface of the pebax.The device was connected to the carto 3 system and no errors were observed.The force feature of the device was evaluated and the force values and the vector were observed within specifications.No force issues were observed.A manufacturing record evaluation was performed for the finished device batch number, and no internal actions were identified.The reddish material inside the pebax could be related to the force issue and foreign material reported by the customer; therefore, the customer complaint was confirmed.The root cause of the pebax damage could be related to the manipulation of the device during the procedure; however, this cannot be conclusively determined.The instruction for use states: do not scrub or twist the tip electrode as damage to the tip electrode bond may occur and loosen the tip electrode.As part of j&j medtech¿s quality process, all devices are manufactured, inspected, and released to approved specifications.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster inc., or its employees that the report constitutes an admission that the product, biosense webster inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's ref.# (b)(4).
 
Event or Problem Description
It was reported that a patient underwent an atrial fibrillation (afib) ablation procedure with a qdot micro for which biosense webster¿s product analysis lab (pal) identified a hole on the pebax.It was initially reported by the customer that after inserting the catheter into the patient's body and immediately after ablation, the contact force suddenly increased.When the catheter was removed from the patient¿s body for inspection, blood was present inside the catheter.There was no difficulty maneuvering the catheter during the withdrawal.The catheter was replaced.The procedure was completed without patient's consequence.The customer's reported force issue is not considered to be mdr reportable since the potential risk that it could cause or contribute to a serious injury or death to the operator or patient is remote.On 16-feb-2026, the bwi pal revealed that a visual inspection of the returned device found a hole in the pebax and a reddish material inside the pebax.These findings were reviewed and assessed the issue of a ¿hole¿ in the pebax as an mdr reportable malfunction since the integrity of the device has been compromised.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QDOT MICRO
Common Device Name
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
henry boden
31 technology dr
irvine, CA 92618
9497882078
MDR Report Key24554398
Report Number2029046-2026-00765
Device Sequence Number13128247
Product Code OAE
UDI-Device Identifier10846835016758
UDI-Public(01)10846835016758(11)250717(17)280716(10)31698644L
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P210027
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberD139505
Device Lot Number31698644L
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/26/2026
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/16/2026
Initial Report FDA Received Date03/09/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/17/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-