Device evaluation details: the qdot micro device was returned to johnson & johnson medtech (j&j medtech) for evaluation.A visual inspection and screening test of the returned device were performed following j&j medtech procedures.Visual analysis revealed a reddish material and a hole in the surface of the pebax.The device was connected to the carto 3 system and no errors were observed.The force feature of the device was evaluated and the force values and the vector were observed within specifications.No force issues were observed.A manufacturing record evaluation was performed for the finished device batch number, and no internal actions were identified.The reddish material inside the pebax could be related to the force issue and foreign material reported by the customer; therefore, the customer complaint was confirmed.The root cause of the pebax damage could be related to the manipulation of the device during the procedure; however, this cannot be conclusively determined.The instruction for use states: do not scrub or twist the tip electrode as damage to the tip electrode bond may occur and loosen the tip electrode.As part of j&j medtech¿s quality process, all devices are manufactured, inspected, and released to approved specifications.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster inc., or its employees that the report constitutes an admission that the product, biosense webster inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's ref.# (b)(4).
|
|
It was reported that a patient underwent an atrial fibrillation (afib) ablation procedure with a qdot micro for which biosense webster¿s product analysis lab (pal) identified a hole on the pebax.It was initially reported by the customer that after inserting the catheter into the patient's body and immediately after ablation, the contact force suddenly increased.When the catheter was removed from the patient¿s body for inspection, blood was present inside the catheter.There was no difficulty maneuvering the catheter during the withdrawal.The catheter was replaced.The procedure was completed without patient's consequence.The customer's reported force issue is not considered to be mdr reportable since the potential risk that it could cause or contribute to a serious injury or death to the operator or patient is remote.On 16-feb-2026, the bwi pal revealed that a visual inspection of the returned device found a hole in the pebax and a reddish material inside the pebax.These findings were reviewed and assessed the issue of a ¿hole¿ in the pebax as an mdr reportable malfunction since the integrity of the device has been compromised.
|