| Model Number |
EM800 |
| Medical Device Problem Code |
No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/27/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the motor was not working properly.It was reported that there was no patient involvement.Additional information was received stating that bearing was damaged due to corrosion inside the equipment found during evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis # (b)(4): evaluation determined that the bearing was damaged due to corrosion inside the equipment.It was also noted that motor's failure.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|