• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. TRILOGY 100; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. TRILOGY 100; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number LA1054096
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/25/2025
Type of Reportable Event Malfunction
Event or Problem Description
The trilogy100 ventilator was returned to the third-party service center in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.There was no allegation of serious or permanent harm or injury.The device was not in use on a patient at the time of the occurrence.The device was returned to the third-party service center for investigation.During evaluation, evidence of (foam degradation inside the assembly, deteriorated foam) was present.There were no repairs performed and the device was scrapped.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY 100
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key24556769
Report Number2518422-2026-008709
Device Sequence Number15807992
Product Code CBK
UDI-Device Identifier00606959022836
UDI-Public(01)00606959022836
Combination Product (Y/N)N
Initial Reporter CountryAR
PMA/510(K) Number
K083526
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 03/10/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberLA1054096
Device Catalogue NumberLA1054096
Device Lot Number120217
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/30/2025
Initial Report FDA Received Date03/10/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/21/2012
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1956-2021
Patient Sequence Number1
Patient SexUnknown
-
-