• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VANTIVE US HEALTHCARE LLC MINICAP; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VANTIVE US HEALTHCARE LLC MINICAP; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE Back to Search Results
Catalog Number 5C4482
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/02/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that there was a separation between the female connector and main body of the minicap transfer set.This occurred during use of the device for peritoneal dialysis therapy.The transfer set was replaced because of the issue.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Additional Manufacturer Narrative
E1: initial reporter postal code:(b)(6).The actual device was not available; however, three (3) photographs of the sample were provided for evaluation.The photographs were reviewed and showed a separation between the female connector and main body.The reported condition was verified.The cause of the separation was due to an inadequate solvent bond between the female connector, insert chip, and main body.A nonconformance has been opened to address this issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MINICAP
Common Device Name
SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
Manufacturer (Section D)
VANTIVE US HEALTHCARE LLC
round lake IL
Manufacturer (Section G)
VANTIVE HEALTHCARE - MOUNTAIN HOME
1900 north highway 201
mountain home AR 72653
Manufacturer Contact
510 lake cook rd
deerfield, il, IL 60015
8153072966
MDR Report Key24557819
Report Number2314912-2026-00161
Device Sequence Number8860070
Product Code KDJ
UDI-Device Identifier00085412007731
UDI-Public(01)00085412007731(10)H23L27093(17)281227
Combination Product (Y/N)N
Initial Reporter CountrySF
PMA/510(K) Number
K152675
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 03/10/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number5C4482
Device Lot NumberH23L27093
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/09/2026
Initial Report FDA Received Date03/10/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/27/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-