| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/02/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that there was a separation between the female connector and main body of the minicap transfer set.This occurred during use of the device for peritoneal dialysis therapy.The transfer set was replaced because of the issue.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
E1: initial reporter postal code:(b)(6).The actual device was not available; however, three (3) photographs of the sample were provided for evaluation.The photographs were reviewed and showed a separation between the female connector and main body.The reported condition was verified.The cause of the separation was due to an inadequate solvent bond between the female connector, insert chip, and main body.A nonconformance has been opened to address this issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|