| Catalog Number |
IAB-06840-U |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/09/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported "during the catheter was inserted into the patient, the balloon ruptured.Change the new catheter".No patient injury or consequence reported.The patient's current condition is reported as "fine".
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The serial number on the returned sample is (b)(6).Returned for investigation was a 40cc 7.5fr ultraflex intra-aortic balloon catheter (iabc) without the original packaging.The sample was returned in a cardboard box and was in a sealed bio-hazard bag.Upon return, the one-way valve was tethered to the short driveline tubing.The bladder was unwrapped.Dried blood was noted on the exterior surfaces of the returned sample.No blood was noted within the helium pathway.No visual damage or abnormalities were noted to the returned sample.The one-way valve was tested and passed.A vacuum was pulled on the one-way valve, and it held for at least 1 minute according to quality system document.The iabc central lumen was successfully aspi-rated and flushed using a 60cc lab-inventory syringe.No abnormalities or debris were noted.The iabc was leak tested in accordance with testing methods from manufacturing procedure.No leaks were detected.Full inflation was achieved.The device passed the leak test.The iabc was connected to an iabp with a lab inventory 40cc inflation driveline tubing.The iabc was inserted into the "t" tube and 100mmhg backpressure was applied.The iabc was pumped for a minimum of 15 minutes.There were no alarms triggered.The bladder inflated and deflated completely with each beat.The balloon pressure waveform (bpw) was normal.A lab inventory 0.025in guidewire was back loaded through the iabc distal tip.No resistance was noted; the guidewire was able to advance through the central lumen.No blood or debris was noted.A device history record (dhr) review was conducted for the lot number with no relevant findings.The device passed all manufacturing specifications prior to release.The reported complaint that the "balloon ruptured" is not confirmed.During the investigation, the returned iab catheter was fully intact.No leaks were detected.The returned device passed visual and functional test specifications.Based on a review of the device history record (dhr), the product met specification upon release.The root cause of the complaint is undetermined.No further action required at this time.
|
| |
|
Event or Problem Description
|
|
It was reported "during the catheter was inserted into the patient, the balloon ruptured.Change the new catheter".No patient injury or consequence reported.The patient's current condition is reported as "fine".
|
| |
|
Search Alerts/Recalls
|