• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. TRADE ITEM NAME BIPAP PRO BIFLEX; VENTILATOR NON-CONTINUOUS (RESPIRATOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. TRADE ITEM NAME BIPAP PRO BIFLEX; VENTILATOR NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number DS660HS
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/05/2026
Type of Reportable Event Malfunction
Event or Problem Description
A bipap pro biflex device was returned to a third-party service center for service as part of the sound abatement foam recall process.During the evaluation of the device, the service technician observed visible foam particles inside the blower kit and object code indicates the blower contributed to the foam degradation.Additionally, device had dust fail contamination.The evaluation of the device data identified unrelated error codes.These error codes are consistent with normal device operation and expected use conditions and are not considered reportable under medical device reporting requirements.No evidence was identified to suggest that these error codes contributed to or caused the reported event.The device was scrapped by the service center after the evaluation was completed.This event is reportable under fda 21 cfr part 803 as it meets the criteria for a serious injury and/or product malfunction.
 
Additional Manufacturer Narrative
Updated the 510k.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRADE ITEM NAME BIPAP PRO BIFLEX
Common Device Name
VENTILATOR NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key24559573
Report Number2518422-2026-008876
Device Sequence Number10114677
Product Code BZD
UDI-Device Identifier00606959005952
UDI-Public(01)00606959005952
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K131982
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation 505
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 03/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberDS660HS
Device Catalogue NumberDS660HS
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/05/2026
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date03/10/2026
Supplement Report FDA Received Date03/17/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/10/2013
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1974-2021
Patient Sequence Number1
Patient SexUnknown
-
-