MEDTRONIC SINGAPORE OPERATIONS ATTESTA L DR MRI SURESCAN; PULSE GENERATOR, PERMANENT, IMPLANTABLE Back to Search Results
Model Number ATDRL1
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Chest Pain (1776); Fall (1848); Palpitations (2467); Unspecified Tissue Injury (4559)
Date of Event 05/01/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was further reported that the patient had a couple of falls, one of which was around the time the rv lead issues began.One of the falls the patient fell on their left shoulder resulting in rotator cuff tearing.It was also noted that the patient was symptomatic.The rv lead further exhibited loss of capture.The ra lead exhibited intermittent capture.No further patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Corrected h6: fdc/fdm/fdr codes: clinical data review was not required.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
B3: date is approximate.Month and year are confirmed valid.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that the patient experienced chest pain and palpitations and was noted to have fallen.It was also reported that the right ventricular (rv) lead exhibited noise, variable and high thresholds, and rising, out-of-range (oor) high impedance.Additionally, the right atrial (ra) lead was reported to exhibit noise, variable, rising, and high thresholds, rising and oor high impedance, and low impedance.The implantable pulse generator (ipg) system was explanted and replaced.No further patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product event summary: the device was returned and analyzed.Returned product analysis was performed and no anomalies were found.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTESTA L DR MRI SURESCAN
Common Device Name
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
MEDTRONIC SINGAPORE OPERATIONS
49 changi south avenue 2
nasaco tech centre
singapore 48605 6
SN  486056
Manufacturer (Section G)
MEDTRONIC SINGAPORE OPERATIONS
49 changi south avenue 2
nasaco tech centre
singapore 48605 6
SN   486056
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112-4391
7635269277
MDR Report Key24559894
Report Number3008973940-2026-01685
Device Sequence Number9593682
Product Code NVZ
UDI-Device Identifier00763000316815
UDI-Public(01)00763000316815(17)230514
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P980035
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/16/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date05/14/2023
Device Model NumberATDRL1
Device Catalogue NumberATDRL1
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/03/2026
Supplement Date Received by Manufacturer03/11/2026
03/25/2026
Initial Report FDA Received Date03/10/2026
Supplement Report FDA Received Date03/25/2026
04/16/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/25/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
5076-58 LEAD, 5076-45 LEAD
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age75 YR
Patient SexMale
-
-