COVIDIEN GIA; STAPLER, SURGICAL Back to Search Results
Model Number GIA8038S
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/02/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
D10.Concomitant products: gia8038l gia 80-3.8sgl use loadingunit - (lot#p2m0406s).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
According to the reporter, during an open colon cancer surgery, during intestinal side-to-side anastomosis, during firing with the reloadable linear stapler, the black push rod took a lot of force to push it even a little bit.After that, a new reload was loaded instead, but the black push rod was very difficult to push.Another device from a different manufacturer was used to resolve the issue.There was no patient injury.
 
Additional Manufacturer Narrative
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device and a photo were available for evaluation.Visual inspection noted that the interlock was engaged.The instrument noted no abnormalities.The single use loading unit (sulu) was partially applied.Microscopic inspection of the knife blade displayed no damage.It was reported that the instrument did not fire.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.This issue may occur if the firing stroke is not complete and the firing knob is not fully retracted after firing or if the instrument is applied against an excessive amount of tissue during the clinical application.If the firing knob is not fully retracted, difficulty in unloading the loading unit may be experienced, and may prevent further loading from occurring.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: to remove the sulu, separate the instrument halves.Holding the cartridge-half of the instrument, grasp the proximal end of the sulu tabs and pull up and out to remove the sulu from the instrument.To place a new sulu into the instrument, hold the sulu by the proximal end tabs and insert into the cartridge fork at a 30 to 45 degree angle from the distal end down until the unit snaps into place.Remove shipping wedge after sulu is fully loaded.Sulu will ¿float¿ up and down in final loaded position.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GIA
Common Device Name
STAPLER, SURGICAL
Manufacturer (Section D)
COVIDIEN
sabanetas industrial park
ponce PR 00716 4401
Manufacturer (Section G)
COVIDIEN
sabanetas industrial park
ponce PR 00716 4401
Manufacturer Contact
mike bocchino
8200 coral sea st ne
mounds view, MN 55112-4391
8006464633
MDR Report Key24560697
Report Number2647580-2026-00635
Device Sequence Number8828920
Product Code GAG
UDI-Device Identifier10884523007268
UDI-Public(01)10884523007268(17)270331(10)P2D0562S(20)01
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K221013
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberGIA8038S
Device Catalogue NumberGIA8038S
Device Lot NumberP2D0562S
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/03/2026
Supplement Date Received by Manufacturer05/28/2026
Initial Report FDA Received Date03/10/2026
Supplement Report FDA Received Date06/11/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/26/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11.
Patient SexUnknown
-
-