A remstar procflex+ device was returned to a third-party service center for service as part of the sound abatement foam recall process.During the evaluation of the device at the third-party service center, the device was visually inspected, and evidence of visible foam particles was found inside the blower kit and foam degradation was also found.Additionally, the service technician also noted dust contamination, and the evaluation of the device data identified unrelated error codes.These error codes are consistent with normal device operation and expected use conditions and are not considered reportable under medical device reporting requirements.No evidence was identified to suggest that these error codes contributed to or caused the reported event.The device was scrapped by the service center after the evaluation was completed.
|