MAUDE Adverse Event Report: ZIMVIE US CORP LLC OSSEOTITE® TAPERED CERTAIN® PREVAIL® IMPLANT 4/3 X 15MM; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMVIE US CORP LLC OSSEOTITE® TAPERED CERTAIN® PREVAIL® IMPLANT 4/3 X 15MM; DENTAL IMPLANT Back to Search Results
Catalog Number XIITP4315
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Osteolysis (2377); Swelling/ Edema (4577)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).A4: weight unknown / not provided.B3: date of event unknown / not provided.A summary investigation has been completed for bone loss and infection events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for bone loss and infection have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of manufacturing or design defects that might lead to bone loss and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
 
Event or Problem Description
It was reported that the implant at tooth site #7 was placed immediately after the extraction of another implant.Excellent paring and stability was noted and a temporary bridge was placed.The patient developed gingival swelling three months later and the implant was removed.Symptoms as a result of the event: inflammation, pain and swelling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OSSEOTITE® TAPERED CERTAIN® PREVAIL® IMPLANT 4/3 X 15MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ZIMVIE US CORP LLC
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZIMVIE US CORP LLC
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
mark mashburn
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key24562561
Report Number0001038806-2026-01281
Device Sequence Number8828982
Product Code DZE
UDI-Device Identifier00844868022145
UDI-Public(01)00844868022145(17)290917(10)2120018774
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K130949
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/10/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue NumberXIITP4315
Device Lot Number2120018774
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/20/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date03/10/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/18/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age61 YR
Patient SexMale
Patient EthnicityNon Hispanic
-
-