On (b)(6) 2025, the healthcare provider (hcp) injected 1ml of evolysse form into a patient's lips.On (b)(6) 2026, the patient noticed some filler oozing from her bottom lip, though she denied manipulating the area herself.The hcp recommended she apply ice to the site intermittently.As of (b)(6) 2026, the patient was out of the country and unable to attend an office visit.On (b)(6) 2026, the hcp contacted the patient again to arrange a follow-up.The patient reported that prior to leaving the country, the filler had oozed from her bottom lip, causing a burning sensation and swelling.While abroad, she was prescribed a five-day course of antibiotics due to concerns about a localized infection.Currently, the patient has a hardened lump on one side of her bottom lip.The reporting hcp has not yet evaluated the patient in person.
|
|
The filler was injected into the patient and is not available for return.The batch release form was reviewed and the device met all specifications prior to release.This is a known potential adverse event addressed in the product labeling.As with all transcutaneous procedures, soft tissue filler implantation carries a risk of infection.This is typically caused by various bacteria entering through a disrupted skin barrier.Additionally, poor aseptic technique, inadequate post-care, or intraoral injections often contribute to this issue.Treatment with antibiotics usually resolves symptoms fully without lasting effects.The product was used off-label in the lip region.We cannot rule out that this may have contributed to the reported event.(b)(4).
|