| Model Number |
8575KA |
| Medical Device Problem Code |
Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device in question was returned to manufacturer.This information is reflected in section d9.The evaluation is anticipated, but not yet begun.The investigation will be performed by a designated karl storz employee.The event is filed under internal karl storz complaint id: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that there was observed a gear malfunction.No patient involvement reported no negative impact in state of health reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the investigation of the returned product revealed chip formation at the thread, which confirmed the customer complaint.The event is filed under internal karl storz complaint id: (b)(4).
|
| |
|