ABBOTT VASCULAR INC. HI-TORQUE COMMAND¿; WIRE, GUIDE, CATHETER Back to Search Results
Model Number 1013753
Medical Device Problem Codes Peeled/Delaminated (1454); Difficult to Remove (1528); Improper or Incorrect Procedure or Method (2017); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/19/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was not returned for analysis.Production record and corrective and preventative actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a query of the complaint handling database for the reported lot revealed there is no indication of a lot specific issue.It was reported the wire was removed with force.It should be noted that ifu, ht guide wires instructions for use states: "carefully observe the instructions under "do not" and "do" below.Failure to do so may result in vessel trauma, guide wire damage, guide wire tip separation, or stent damage.If resistance is observed at any time, determine the cause under fluoroscopy and take remedial action as needed.Do not: push, auger, withdraw or torque a guide wire that meets resistance.Torque a guide wire if the tip becomes entrapped within the vasculature.The ifu deviation likely did not cause the reported issue.Based on the information provided, a definitive cause for the reported issue could not be determined.In this case the wire may have been damaged during delivery, during delivery of the cardiomems; however, this cannot be confirmed.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.The additional device mentioned in b5 will be filed under a separate medwatch report.
 
Event or Problem Description
It was reported the procedure was to implant a cardiomems sensor in a pulmonary artery.Two command 18 lt guide wires were used however during removal, resistance was met with the cardiomems delivery catheter.Force was applied in the attempts to remove the guide wire and once removed from the anatomy, it was noted the polymer coating on the guide wires were peeling [degloved] and twisted.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HI-TORQUE COMMAND¿
Common Device Name
WIRE, GUIDE, CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR INC.
3200 lakeside drive
santa clara CA 95054 2807
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key24567478
Report Number2024168-2026-00983
Device Sequence Number15595711
Product Code DQX
UDI-Device Identifier08717648211980
UDI-Public(01)08717648211980(11)250217(17)270131(10)5021761
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K172073
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1013753
Device Catalogue Number1013753
Device Lot Number5021761
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/20/2026
Initial Report FDA Received Date03/11/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/17/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CARDIOMEMS DELIVERY CATHETER
Patient SexUnknown
-
-