|
H3: product analysis of part# 5200-1001, lot# y210906 visual inspection confirmed the tip of the inserter has been bent and the inner spring is missing.The damage to the inserter tip is consistent with bend stress overload.This regulatory report is being submitted due to retrospective review.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|