DEPUY SYNTHES PRODUCTS LLC HIGH SPEED ELEC G1 HANDPIECE AIR COOLED; DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE Back to Search Results
Catalog Number EG1A
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11 additional narrative: investigation summary: a visual and functional assessment was performed on the device: during the service evaluation the following failures were identified: functional: heat.Conclusion: failure reported is confirmed.Root cause: improper maintenance.Action: repair.
 
Event or Problem Description
It was reported that during service and evaluation; it was determined that the handpiece device had heat.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention, or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2026.All available information has been disclosed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH SPEED ELEC G1 HANDPIECE AIR COOLED
Common Device Name
DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
elba bello
4500 riverside drive
palm beach gardens, FL 33410
9083863534
MDR Report Key24568021
Report Number1045834-2026-00065
Device Sequence Number18231568
Product Code ERL
UDI-Device Identifier00845384017134
UDI-Public(01)00845384017134
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K133604
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberEG1A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/27/2026
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/05/2026
Initial Report FDA Received Date03/11/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/29/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-