ZIMMER BIOMET, INC. BIOMET HIP SYSTEM; PROSTHESIS, HIP Back to Search Results
Catalog Number 11-363662
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Joint Dislocation (2374); Joint Laxity (4526)
Date of Event 02/16/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that approximately two and a half years post-implantation, the patient underwent a left hip revision due to instability and recurrent dislocation.All components revised.Attempts have been made and no further information has been provided.
 
Additional Manufacturer Narrative
(b)(4).D10: cat 010000704 lot# 7549541 g7 bonemaster ltd acet shl 54f; cat# 51-117140 lot# 7303049 tprlc 133 mp type1 bm ho 14.0; g2: foreign - event occurred in australia.H6: proposed component code: mechanical (g04) - head.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Event or Problem Description
No further information is available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: d4; h2; h3; h4; h6; h10.Complaint is confirmed.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history record identified no deviations or anomalies during manufacturing.Review of complaint history identified additional similar complaints for the reported item and no additional complaints for the reported part and lot combination.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: posterior superior dislocation of the left total hip arthroplasty.Of note, vertical inclination of the acetabular cup can predispose to dislocation.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOMET HIP SYSTEM
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key24568119
Report Number0001825034-2026-00552
Device Sequence Number21207312
Product Code JDI
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K032396
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue Number11-363662
Device Lot NumberJ7462329
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/16/2026
Supplement Date Received by Manufacturer06/08/2026
Initial Report FDA Received Date03/11/2026
Supplement Report FDA Received Date06/23/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/07/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age64 YR
Patient SexUnknown
Patient Weight293 LB
-
-