|
A remstar pro c flex device was returned to a third party service center for service as part of the sound abatement foam recall process.During the evaluation of the device, the service technician observed visible foam particles inside the blower kit and object code indicates the blower contributed to the foam degradation.Additionally, the device had fluids fail contamination.The evaluation of the device data identified unrelated error codes.These error codes are consistent with normal device operation and expected use conditions and are not considered reportable under medical device reporting requirements.No evidence was identified to suggest that these error codes contributed to or caused there reported event.The device was scrapped by the service center after the evaluation was completed.This event is reportable under fda 21 cfr part 803 as it meets the criteria for a serious injury and/or product malfunction.
|