• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC SOFAMOR DANEK USA, INC. CATALYFT¿ PL EXPANDABLE INTERBODY SYSTEM; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT,

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC SOFAMOR DANEK USA, INC. CATALYFT¿ PL EXPANDABLE INTERBODY SYSTEM; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, Back to Search Results
Model Number 6061000
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/06/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H3: product analysis of part# 6061000, lot # kh21f250.Visual and optical inspection did not reveal any rust on the inner shaft but confirmed the tip of the inserter is worn from repeated use.This regulatory report is being submitted due to retrospective review.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Information was received from regulatory body (rb), user facility via a manufacturer representative regarding a device used for tlif interbody insertion spinal therapy.It was reported that device was rusted and broken. there was no patient involvement reported and no further complications reported regarding the event additional information was received via manufacturer representative that inner portion of the inserter has rusted out.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CATALYFT¿ PL EXPANDABLE INTERBODY SYSTEM
Common Device Name
INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT,
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK USA, INC.
4340 swinea rd
memphis TN 38118
Manufacturer (Section G)
MEDTRONIC SOFAMOR DANEK USA, INC.
4340 swinea rd
memphis TN 38118
Manufacturer Contact
justin ellis
1800 pyramid place
memphis, TN 38132
7635265677
MDR Report Key24569340
Report Number3003120897-2026-00432
Device Sequence Number9561310
Product Code MAX
UDI-Device Identifier00763000253530
UDI-Public(01)00763000253530(11)210719(10)KH21F250
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K210425
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 03/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number6061000
Device Catalogue Number6061000
Device Lot NumberKH21F250
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/17/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/19/2025
Initial Report FDA Received Date03/11/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/19/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-