Model Number M635WU50270
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Bacterial Infection (1735)
Date of Event 02/27/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
B3 date of event - aware date used as event date is unknown.
 
Event or Problem Description
It was reported that there was an infection.A left atrial appendage (laa) closure procedure was being performed.A watchman access system (was) and a 27mm watchman flx laa closure device & delivery system (wds) were selected to be used.The procedure was successfully completed with the closure device implanted in the laa of the patient.There were no patient complications that were reported to have occurred.Two years post procedure it was reported that the patient had been diagnosed with mrsa found on the closure device.
 
Additional Manufacturer Narrative
B3 date of event - aware date used as event date is unknown.With all the available information, boston scientific (bsc) concludes: device technical analysis: the watchman flx? remains implanted; therefore, returned device analysis could not be completed.Device history record review: a review was completed of the device history records (dhr) for the watchman flx?, part # m635wu50270 batch # 0031328875.The device was processed through all normal operational conditions and passed all acceptance activities.The dhr review did not identify any deviations or non-conformances that could have contributed to the event.Labeling review: there was no evidence to suggest that the device was used in a manner inconsistent with the labeling.Watchman flx? instructions for use (ifu) lists potential complications, risks and adverse events that may be associated with the use of this device which include infection (serious injury/illness/impairment).Risk review: a risk review was completed and confirmed that the event of infection was defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Investigation conclusion: based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of known inherent risk of device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WATCHMAN FLX?
Common Device Name
SYSTEM, APPENDAGE CLOSURE, LEFT ATRIAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key24570317
Report Number2124215-2026-13590
Device Sequence Number21222659
Product Code NGV
UDI-Device Identifier08714729860501
UDI-Public(01)08714729860501(17)260328(10)31328875
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P130013
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/28/2026
Device Model NumberM635WU50270
Device Catalogue NumberM635WU50270
Device Lot Number0031328875
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/27/2026
Supplement Date Received by Manufacturer04/24/2026
Initial Report FDA Received Date03/11/2026
Supplement Report FDA Received Date04/29/2026
Date Device Manufactured03/29/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age76 YR
Patient SexMale
-
-