| Catalog Number |
381823 |
| Medical Device Problem Code |
Retraction Problem (1536)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/28/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H.3.A follow up mdr will be submitted if additional information, a device evaluation, or a device history review is completed.
|
| |
|
Event or Problem Description
|
|
It was reported that the needle retracted prematurely.Spring decompressed.Impact caused to the patient / end user: none.Date of event: 28.09.2025.Additional information received on 07-mar-2026.When was the problem/defect detected: before, during or after use? before.
|
| |
|
Additional Manufacturer Narrative
|
|
The complaint of a decompressed spring, which is associated with premature needle retraction, was confirmed from the photograph that was provided for investigation.The photos showed one 22g insyte autoguard device in what appeared to be sealed unit packaging.The needle appeared to be fully retracted within the safety shield based on the position of the white flow control plug in relation to the blue catheter adapter.As the sample appeared to be unopened, the safety mechanism was likely activated prior to customer use.The appropriate manufacturing personnel were notified of this complaint.A review of our risk management documentation was performed and indicates that the potential risk of the reported event was assessed appropriately.Complaints received about this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the data collected for identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
No new information.
|
| |
|