Model Number QRP6
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/21/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation confirmed the reported issue.Bobbin assembly was replaced and unit was confirmed to operate correctly.This event was originally reported in a summary report 3006073153-2025-00059, which was submitted in relation to the affected devices from the voluntary recalled performed by spectrum medical ltd (recall number z-0224-2025).Fda has since requested that each event for the devices related to the voluntary recall be individually submitted, resulting in the creation and submission of this report.
 
Event or Problem Description
User reported that one of the rollers had bobbins that would loosen when the lever was in the locked position.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUANTUM ROLLER PUMP - 6"
Common Device Name
ROLLER PUMP
Manufacturer (Section D)
SPECTRUM MEDICAL LTD
ashville business park
commerce road
gloucester, gloucestershire gl2 9sl
UK 
Manufacturer (Section G)
SPECTRUM MEDICAL LTD
ashville business par
commerce road
gloucester, gloucestershire gl2 9sl
UK  
Manufacturer Contact
raffaella tommasini
ashville business park
commerce road
gloucester, gloucestershire gl2 9sl 
UK  
MDR Report Key24573892
Report Number3006073153-2026-00063
Device Sequence Number9232560
Product Code DTQ
Combination Product (Y/N)Y
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K173834
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 03/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberQRP6
Device Catalogue Number51-000013-00
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/21/2023
Initial Report FDA Received Date03/11/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/08/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0224-2025
Patient Sequence Number1
Patient SexUnknown
-
-