DEPUY MITEK LLC US LUPINE BR DS W/ORTHCRD; SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE
|
Back to Search Results |
|
| Catalog Number |
210712 |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/05/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.E1: the reporter's complete facility address was not provided.H11: as of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated, and no conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
It was reported that during a repair of anterior talofibular ligament surgery, the lupine br ds w/orthcrd anchor was pulled out.Another device was used to complete the surgery.There was a five minute delay.There were no adverse consequences to the patient.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: the product was not returned to depuy synthes, however a photo was provided for evaluation.Visual inspection of the returned photo found that the anchor was detached from the inserter and it was not visible in the image.The shaft and sutures have no structural anomalies.Without the anchor, it is not possible to confirm the reported condition.No other anomalies were noted.The overall complaint was not confirmed as the observed condition of the lupine br ds w/orthcrd would have not contributed to the complained device issue.Based on the findings, the investigation could not draw any conclusions and no potential cause about the reported event can be established due to the insufficient evidence provided.It has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
|
| |
|
Search Alerts/Recalls
|
|
|