• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COCHLEAR BONE ANCHORED SOLUTIONS AB BIA400 IMPLANT 4MM W ABUTMENT 8MM; COCHLEAR BAHA CONNECT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COCHLEAR BONE ANCHORED SOLUTIONS AB BIA400 IMPLANT 4MM W ABUTMENT 8MM; COCHLEAR BAHA CONNECT SYSTEM Back to Search Results
Model Number 93330
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code Skin Infection (4544)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reported adverse event is associated with a returned device; however, the provided clinical information was reviewed by the manufacturer and no specific device analysis is deemed necessary at this time.Previous product examinations have not showed any relationship between a product geometrical deviation and the reported clinical complication.Additionally, there are no indications that a product failure has contributed to the reported issue.Adverse skin reactions around the baha abutment, ranging from slight redness to infected tissue, are common complications with baha implants, and can occur at any time following implantation.The report frequency for these complications is being monitored under cbas complaint and medical device reporting data monitoring plan and the status is updated on a monthly basis.This event is added to this monitoring.
 
Event or Problem Description
Per the clinic, the patient developed an infection and subsequently the device was explanted on (b)(6) 2026.There are no plans to reimplant the patient with a new device as of the date of this report.Additional information has been requested but has not been made available as of the date of this report.
 
Additional Manufacturer Narrative
Per the clinic, the patient was treated with oral antibiotics on (b)(6) 2025, administered three times daily for 5 days.The abutment was also reportedly removed on the same date.Subsequently, on (b)(6) 2025, the patient was prescribed antiseptic treatment and topical antibiotics, to be applied twice daily for 14 days.Following the appointment on (b)(6) 2025, the patient received an additional course of oral antibiotics (specific duration not reported).During the explant surgery on (b)(6), 2026, the surgeon encountered a small maxillofacial screw, which was drilled out and removed.No infection was reported following the explant surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIA400 IMPLANT 4MM W ABUTMENT 8MM
Common Device Name
COCHLEAR BAHA CONNECT SYSTEM
Manufacturer (Section D)
COCHLEAR BONE ANCHORED SOLUTIONS AB
konstruktionsvägen 14
mölnlycke, 435 3 3
SW  435 33
Manufacturer (Section G)
COCHLEAR BONE ANCHORED SOLUTIONS AB
konstruktionsvägen 14
mölnlycke, 435 3 3
SW   435 33
Manufacturer Contact
nabila zahari
unit ug-1, vertical podium
no. 8 jalan kerinchi,
kuala lumpur, wilayah persekutuan 59200
MY   59200
MDR Report Key24577350
Report Number6000034-2026-00882
Device Sequence Number13118212
Product Code MAH
UDI-Device Identifier09321502022705
UDI-Public(01)09321502022705(10)COH1378762(17)250324
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K121317
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2026
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date03/24/2025
Device Model Number93330
Device Catalogue Number93330
Device Lot NumberCOH1378762
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/11/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 02/18/2026
Supplement Date Received by Manufacturer04/07/2026
Initial Report FDA Received Date03/12/2026
Supplement Report FDA Received Date04/29/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-