| Model Number |
6246011 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Product Quality Problem (1506)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/27/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Information was received from a manufacturer representative regarding an event which occurred during servicing.It was reported that new set came with inserters and inner shaft which don't thread in fully, it was noticed that button of the driver sticks and it appears tip of inner shaft is bent slightly and it was noticed that extender tab was bent. there was no patient involvement reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: product analysis of part# 6246011, lot # em23g002 functional inspection confirmed the inner threads are worn from repeated use.This regulatory report is being submitted due to retrospective review.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|