• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMVIE US CORP LLC CERTAIN® TITANIUM HEXED SCREW; DENTAL SCREW

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMVIE US CORP LLC CERTAIN® TITANIUM HEXED SCREW; DENTAL SCREW Back to Search Results
Catalog Number IUNIHT
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).D10: concomitant medical product: ilrght, certain® titanium large hexed screw, lot: 1299289.
 
Event or Problem Description
It was reported that the screws were loosened and did not hold the crowns properly, and the screws were removed.No other screws were used in exchange.
 
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).The following fields have been updated: b4: date of this report b5: describe event or problem g3: date received by manufacturer g6: type of report h1: type of reportable event h2: follow up type h3: device evaluated by manufacturer h6: adverse event problem h11: additional narrative zimvie received one (1) iuniht, (certain titanium hexed screw) for evaluation.Visual evaluation and a functional test were performed, the screw shows signs use.The screw threaded into in-house abutment and implant as intended.Measurements match drawing and engaged thread ring.Unable to confirm loosening with all the available information.Dhr review was completed for the subject lot number 1258670.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 1258670 for similar events and no other complaint was identified.Review completed utilizing keywords: ¿dental : functional : loosening¿ based on the investigation and risk management file review, the most likely root cause determined from the investigation was the abutment screw was not torqued to specification.Therefore, based on the available information, a device malfunction could not be verified.The screw threaded into in-house implant as intended.The reported loosening event was non-verifiable with all the available information.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CERTAIN® TITANIUM HEXED SCREW
Common Device Name
DENTAL SCREW
Manufacturer (Section D)
ZIMVIE US CORP LLC
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZIMVIE US CORP LLC
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
mark mashburn
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key24578130
Report Number0001038806-2026-01330
Device Sequence Number9435453
Product Code NHA
UDI-Device Identifier00844868008545
UDI-Public(01)00844868008545(17)270929(10)1258670
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K072642
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 05/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue NumberIUNIHT
Device Lot Number1258670
Was Device Available for Evaluation? No
Date Returned to Manufacturer02/26/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date03/12/2026
Supplement Report FDA Received Date05/11/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/30/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DENTAL SCREW SEE H10 NARRATIVE.
Patient SexUnknown
Patient EthnicityNon Hispanic
-
-