| Catalog Number |
138104 |
| Medical Device Problem Code |
Delivered as Unsterile Product (1421)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Received 10 of 138104 in original packaging.Lot number was verified.Performed a visual inspection of the devices, there were obvious signs of a breach on six of the devices.Performed a functional inspection on the remaining four devices, they were dye leak tested which indicated that all four of the packages had an insufficient heat seal as there was a leak.Root cause cannot be determined, however, based upon evaluation of the device; a possible cause of this event could be the device was encroaching the seal.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution was found to have met all specifications prior to shipment.(b)(4).Per the instructions for use, the user is advised the following: inspect and test each device before each use.We will continue to monitor per regulatory requirements to help ensure patient safety.
|
| |
|
Event or Problem Description
|
|
During incoming inspection, the distributor rejected this device, 138104 for an insufficient heatseal.There was no contact with the patient as this was found during incoming inspection.This report is being raised due to the reported malfunction with potential for injury upon reoccurrence.
|
| |
|
Search Alerts/Recalls
|