During incoming inspection, the distributor rejected this device, 138114a, for an insufficient heatseal.There was no contact with the patient as this was found during incoming inspection.This will be reported as a malfunction with potential for injury upon reoccurrence.
|
|
Received three of 138114a in original packaging.Lot number was verified.Performed a visual inspection of the device, there were obvious signs of a breach on two of the devices.Performed a functional inspection on the third device, it was dye leak tested per tm-10-217 rev.F which indicated that the last package had an insufficient heat seal as there was a leak.Root cause cannot be determined, however, based upon evaluation of the device and the ifu; a possible cause of this event could be the package was improperly sealed.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution was found to have met all specifications prior to shipment.A two-year lot history review shows a total of (b)(4) devices for this lot number and failure mode.(b)(4).Per the instructions for use, the user is advised that sterility guaranteed unless package has been opened, broken, or damaged.Inspect and test each device before each use.We will continue to monitor per regulatory requirements to help ensure patient safety.
|