• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NATUS MEDICAL INCORPORATED NT821731C, EDS 3, GEN LL, NO CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NATUS MEDICAL INCORPORATED NT821731C, EDS 3, GEN LL, NO CATHETER Back to Search Results
Model Number NT821731C
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/19/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Initial report ref to natus complaint# (b)(4).(b)(4) rev 13 risk analysis spreadsheet - eds 3 external drainage system.Hazard 4.36 - stopcock valve unable to turn / rotate and/or handle breakage.Effect (harm): delay in patient treatment, over / under drainage of csf and infection.Residual risk: medium.The hazards identified have been reduced as far as possible, and the benefit of use of the entire product outweighs the risks identified.Reference to capa005781.Further investigation to be carried out.
 
Event or Problem Description
Nt821731c - evd stopcock break.It was the stopcock at the transducer site.No harm to patient.
 
Additional Manufacturer Narrative
Follow up report 001 ref to natus complaint# (b)(4).Sterile date of product: (b)(6) 2025.Device history record review: unit has passed all the required tests.No manufacturing defects or non-conformances observed.A review of (b)(4) medical device hazard analysis (rev 13), identifies the hazard situation as "4.36 - stopcock valve unable to turn or rotate".The risk level is considered moderate.Based on complaint of "stopcock broke." this determination is based on the information received from the customer.Related to capa005781.Root cause/failure investigation: this case had a drainage system with a damaged system stopcock.The stopcock was broken at the lever which is used to turn the stopcock into a closed or open position.The system stopcock material exhibited significant deformation.The breakage of the components indicates that the user turned the stopcocks with excessive torque possibly with a tool instead of by hand.Another possible indication could be that the client used aggressive cleaning agents that degrade the polycarbonate material.The drip chamber stopcock was broken at the wall.There was a buildup of dried blood/bodily fluids at the drip chamber stopcock.Assessment of whether bodily fluid exposure created an increased resistance during stopcock manipulation was conducted.The stopcock exhibited an elevated rotational resistance relative to baseline performance observed in new units.Section "warnings" on page 3 of the eds 3 system ifu (sd208650001rev e) gives details on handling the eds 3 system such as finger tightening components and not using external tools.Failure mode: confirmed / stopcock breakage during use.
 
Event or Problem Description
Nt821731c - evd stopcock break.It was the stopcock at the transducer site.No harm to patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NT821731C, EDS 3, GEN LL, NO CATHETER
Common Device Name
NT821731C, EDS 3, GEN LL, NO CATHETER
Manufacturer (Section D)
NATUS MEDICAL INCORPORATED
5955 pacific center boulevard
san diego CA 92121
Manufacturer Contact
5955 pacific center boulevard
san diego, CA 92121
MDR Report Key24581267
Report Number2023988-2026-00034
Device Sequence Number18422215
Product Code JXG
UDI-Device Identifier00382830050555
UDI-Public00382830050555
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K162437
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberNT821731C
Device Catalogue NumberNT821731C
Device Lot Number118001157058
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 02/20/2026
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date03/12/2026
Supplement Report FDA Received Date04/25/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/06/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age41 YR
Patient SexMale
-
-