Catalog Number H787103028035/A
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/20/2026
Type of Reportable Event Serious Injury
Event or Problem Description
The account alleges that during device removal through the skin tract, the metal component separated from the plastic sleeve and remained under the patient's skin.The retained component was manually retrieved.No patient harm was reported.
 
Additional Manufacturer Narrative
The suspect device is expected to return for evaluation.A follow up will be submitted when the evaluation is complete.
 
Additional Manufacturer Narrative
The suspect device was not returned for evaluation.The complaint could not be confirmed.The root cause could not be determined.A search of the complaint database was performed and no similar complaints for this lot number were found.The device history record was reviewed, and no exception documents were found.Nonconformances of this nature are monitored by the operation team.This complaint will be used to trend for similar complaints.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DURAMAX®
Common Device Name
CATHETER, HEMODIALYSIS, IMPLANTED
Manufacturer (Section D)
MERIT MEDICAL SYSTEMS, INC.
calle, sor juana ines de la cr
roberto de la madrid
tijuana, b.c. 22630
MX  22630
Manufacturer (Section G)
MERIT MEDICAL SYSTEMS, INC.
calle, sor juana ines de la cr
roberto de la madrid
tijuana, b.c. 22630
MX   22630
Manufacturer Contact
adrian cole
1600 merit parkway
south jordan, UT 84095
8012084225
MDR Report Key24583611
Report Number1721504-2026-00914
Device Sequence Number10325713
Product Code MSD
UDI-Device Identifier00884450786130
UDI-Public00884450786130
Combination Product (Y/N)Y
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K101843
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/07/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Catalogue NumberH787103028035/A
Device Lot NumberI3288649
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/20/2026
Supplement Date Received by Manufacturer03/31/2026
Initial Report FDA Received Date03/12/2026
Supplement Report FDA Received Date04/07/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/20/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-