| Catalog Number |
201001040 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/26/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that broaches are broken into two pieces.No additional fragments.All pieces were retrieved.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: it was reported that this facility had multiple version 1 and 2 broaches break and were requesting all of their consigned broaches be swapped out for version 3 to avoid further breakages.Additional information received; both broach breakages were in exactly two pieces.Separate ders were filed for each individual incident, no additional fragments.All pieces were retrieved and returned.The product was returned to depuy synthes for evaluation.Visual inspection of the returned device found nothing indicative of a device nonconformance.No broken areas were identified.Device only displays sings of repetitive use.A functional test was not performed since it was not applicable to the complaint condition.A dimensional inspection was not performed since it was not applicable to the complaint condition.The overall complaint was not confirmed as the observed condition of the actis broach would have not contributed to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.Corrected: h3.
|
| |
|
Search Alerts/Recalls
|