| Model Number |
VTICM5_13.2 (-11.50/1.5) |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Endophthalmitis (1835); Blurred Vision (2137)
|
| Date of Event |
02/02/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Secondary surgical intervention to remove/replace/reposition is identified in the labeling as a known potential adverse event(s) following icl implantation.Inflammation is identified in the labeling as a known potential adverse event following icl implantation.The dfu provides the surgeon instruction for complete ovd removal.Precaution: (6) after inserting this product, aspirate the viscoelastic substance completely from inside the eye.Do not use highly viscous viscoelastic substances that are difficult to aspirate completely.H6: investigation type 4110: a lens work order search was performed.No additional similar complaint type events within associated lots were found.Claim: (b)(4).
|
| |
|
Event or Problem Description
|
|
A facility representative reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced endophthalmitis, hypopyon, and loss of bcva.The lens was explanted, antibiotic medication was administered, and the problem was resolved.Additional information was requested.No further information is available at this time.This file will be re-opened if additional information is provided.
|
| |
|
Additional Manufacturer Narrative
|
|
H6 - type of investigation 3331 - device history record (dhr) review: based on the results of the investigation, all released devices from the associated work order(s), including the suspected device, have been manufactured within established process parameters and there is no indication that the manufacturing and processing of the device contributed to the complaint issue.Claim # (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information
b5 - describe event or problem - a facility representative reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced endophthalmitis, hypopyon, and loss of bcva.Diagnotic culture was performed.The lens was explanted, antibiotic medication was administered, and the problem was resolved.Claim # (b)(4).
|
| |
|