Catalog Number 138107
Medical Device Problem Code Delivered as Unsterile Product (1421)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Received one of 138107 in original packaging.Lot number was verified.Performed a visual inspection of the device, there were no obvious signs of a breach.Performed a functional inspection, the device was dye leak tested which indicated that the package had an insufficient heat seal as there was a leak.Root cause cannot be determined, however, based upon evaluation of the device; a possible cause of this event could be the device was encroaching the seal.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution was found to have met all specifications prior to shipment.A two-year lot history review shows a total of one device for this lot number and failure mode.A two-year review of complaint history revealed there has been a total of 221 complaints, regarding (b)(4) devices, for this device family and failure mode.During this same time frame (b)(4) devices have been manufactured and shipped worldwide.Should all the complaint devices have been found confirmed for this reported failure, the rate of failure would be (b)(4).Per the instructions for use, the user is advised to inspect and test each device before each use.We will continue to monitor per regulatory requirements to help ensure patient safety.
 
Event or Problem Description
This distributor reported that the 138107 device had an insufficient heatseal found during incoming inspection.No patient involvement occurred.This report is being raised due to the reported malfunction with potential for injury upon reoccurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONMED
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
CONMED
525 french rd
utica NY 13502
Manufacturer (Section G)
CONMED UTICA
525 french rd
utica NY 13502
Manufacturer Contact
samantha dewberry
525 french road
utica, NY 13502
3152230184
MDR Report Key24586184
Report Number1320894-2026-00054
Device Sequence Number9385911
Product Code GEI
UDI-Device Identifier30653405003421
UDI-Public(01)30653405003421(17)290127(10)202401291
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K943542
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 03/12/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number138107
Device Lot Number202401291
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/23/2026
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/24/2026
Initial Report FDA Received Date03/12/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/29/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
Patient EthnicityNon Hispanic
-
-