| Model Number |
20000800100011010 |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during biomed testing, the device would not power up.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation for evaluation.During disassembly of the device to determine root cause, the technician observed damage not consistent with typical use.A visual inspection of the device found damage to the power button not consistent with normal wear and tear.The buttons on the device should not be pressed with sharp objects, only pressed with fingers.The on/off gasket and main board were replaced.The device was recertified and returned to the customer.
|
| |
|