An extended investigation has been conducted, which involved a manufacturing review (including 5 years of device history records (dhrs)), previous complaint and corrective and preventive action (capa) investigations conducted for blank screen issues, process failure mode effects analyses (pfmeas), design controls, and design failure mode effects analyses (dfmeas), risk management reports, risk evaluations, and label copy.The investigation did not identify any indication that the product did not meet specification.The returned meter was investigated with retained test strips and a known-good retained battery.Visual inspection was performed on returned meter and no issues were observed.Meter powered on with button depression and test strip insertion.Blank screen was not observed.The complaint was not confirmed.All pertinent information available to abbott diabetes care has been submitted.
|
|
The reporter contacted abbott diabetes care, alleging the meter will not turn on.This complaint is being reported because the reported issue was not resolved with troubleshooting.There was no report of death, serious injury, or mistreatment associated with this event.
|