DEPUY ORTHOPAEDICS INC US LPS CEMENTED STEM 11X100MM STR; LPS AND S-ROM : KNEE STEMS Back to Search Results
Catalog Number 198725111
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 03/02/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the stem was broken postoperatively.The revision surgery was planned to be preformed on (b)(6) 2026.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H11.Additional narrative: added: b5.Corrected: h6 (health effect - impact code).If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
Additional information received: was there any revision surgery performed? it was planned on (b)(6) 2026.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: stem broken.The product was not returned to depuy synthes, however photos were provided for review.The photo investigation revealed that the lps cemented stem 11x100mm str shows a fractured near its middle section.The mode of failure of the device is multi-factorial and consideration must be given to all other potential influences such as surgical process, patient variables (e.G.Activity level and use), anatomical considerations and patient changes over time.Any conclusions from the investigational input provided have to be placed into context with all other relevant factors.As the device was not returned, and as-received condition could not be assessed, a dimensional inspection and document/specification review were not completed.The overall complaint was confirmed as the observed condition of the lps cemented stem 11x100mm str would contribute to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: a manufacturing record evaluation was performed for the finished device j70x28 number, and no non-conformances / manufacturing irregularities were identified during manufacturing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LPS CEMENTED STEM 11X100MM STR
Common Device Name
LPS AND S-ROM : KNEE STEMS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
elba bello
700 orthpaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key24590754
Report Number1818910-2026-04767
Device Sequence Number15835671
Product Code JWH
UDI-Device Identifier10603295079156
UDI-Public(01)10603295079156(17)310228(10)J70X28
Combination Product (Y/N)N
Initial Reporter CountryAU
PMA/510(K) Number
K033959
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number198725111
Device Lot NumberJ70X28
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/02/2026
Supplement Date Received by Manufacturer03/17/2026
04/09/2026
Initial Report FDA Received Date03/13/2026
Supplement Report FDA Received Date03/25/2026
04/10/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/24/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-